Radio-frequency sensing · NQR

NQR-based RF Magnetometer

Designing an atomic radio-frequency magnetometer for nuclear quadrupole resonance detection.

NQR magnetometer project illustration

Project Abstract

At A*STAR Q.InC, I contributed to the first half of a project to build a prototype NQR-based radio-frequency magnetometer for detecting unexploded ordnance and specific crystalline narcotics. By the time I left the project, we had completed the literature review, designed the experimental setup, and assembled it for initial testing.

The sensor is closely related to the QEP 2.0 Atomic Magnetometer, with upgrades to the optics and vapour-cell geometry for RF operation. These include RF and detuning coils, modular sample mounts, and pump-power amplification.

Background And Motivation

Nuclear Quadrupole Resonance (NQR) is a technique in which certain nuclei within molecules have energy levels that split because their quadrupole moments interact with the electric-field gradients created by their local chemical environment. Radio-frequency transitions between these levels have characteristic frequencies that depend on the local chemical bonds and crystal structure.

Experimentally, we can drive and then probe these transitions by applying a radio-frequency magnetic field to the sample. This excites the nuclear spins and induces an oscillating nuclear magnetization. Through careful engineering, we propose to use an atomic magnetometer to measure this magnetization instead of a conventional RF receiver coil, with the aim of improving sensitivity.

We focus on NQR spectroscopy of nitrogen-14 nuclei, which are found in many explosives and crystalline narcotics. Potential applications include portable or installed screening near import facilities, immigration checkpoints, and other sites requiring stringent national security.

Schematic comparison of prolate, spherical, and oblate nuclear shapes
Schematic nuclear shapes illustrate how a non-spherical nuclear charge distribution can have an electric quadrupole moment.
Close view of the RF excitation and detuning coil region inside the magnetometer shielding
Close view of the RF excitation and detuning-coil region inside the magnetometer shielding.
Results And Next Steps

When I left the project to pursue my MSc, the literature review was complete and the experimental setup had been designed and assembled for initial testing. Testing the sensor and RF excitation sequence with reference samples was the next step.

The design review set a sensitivity target below 2 fT/√Hz. Simulations predicted an optimum below 1 fT/√Hz and estimated that a signal-to-noise ratio of 3 could be reached from about 70 g of ammonium nitrate in less than 20 seconds. The first functional test would be to detect ammonium nitrate, then progressively test compounds with higher NQR resonance frequencies.

Reference table of NQR resonance frequencies and relaxation parameters for ammonium nitrate, TNT, PETN, RDX, HMX, cocaine base, and caffeine
Reference NQR resonance frequencies and relaxation parameters for candidate compounds. The listed frequencies indicate a possible progression from the initial ammonium-nitrate test to higher-frequency targets.
Nitrogen-14 NQR resonance-frequency bands for RDX, HMX, PETN, and TNT, with an approximately 5 kHz excitation bandwidth marked
Example nitrogen-14 NQR resonance frequencies for several compounds; the marked 5 kHz bandwidth illustrates the narrow frequency range addressed by an excitation sequence. Source: IAEA conference paper by Zhu.

Vapour Cell, Oven, Magnetic Coils, Cooling, And Modularity

The experiment required custom mechanical supports for its critical components; their functions are described in the Methodology section. We designed and assembled the setup, including the magnetic-coil wiring, a thermally isolating oven, the 20 × 20 × 20 mm potassium vapour cell, and a ceramic RF-coil assembly with a mount designed for active water cooling. The sample holder and surrounding structure were designed to make changing chemical samples straightforward. See the images at the end for 3D renders of the experimental setup.

Pump Optics And Tapered Amplifier

The pump system is more powerful and uses a larger beam than the QEP 2.0 magnetometer. A tapered amplifier boosts the seed laser to approximately 121 mW, with the beam expanded and shaped to illuminate the vapour cell diffusely and relatively uniformly. Other optics set the beam size at the cell, improve coupling into and out of the amplifier, and provide an additional isolator to reduce back reflections.

Electronics

We developed custom electronics for the experiment. I designed an interlock system that monitors pump-laser power and shuts down the amplifier if the seed laser is absent. We also measured the RF coils' inductance in situ to inform the matching network design, reduce unintended parasitic capacitance, and support efficient resonance matching. This matching is important for reaching the higher resonance frequencies of some target compounds, several of which have few or no demonstrated detections in the literature.

Methodology
Top and side views of the NQR radio-frequency atomic magnetometer, showing pump and probe paths, shield, field coils, RF sample coils, sample holder, oven, and vapor cell
Optical, magnetic, and RF layout of the NQR magnetometer. The side view shows the sample, excitation coils, and vapor cell assembly.

Sample Excitation And Atomic Sensing

Sample holder and RF coils: the modular holder positions the sample next to the potassium vapour cell. The top RF coil applies the radio-frequency field that excites the sample's NQR transition. The bottom RF coil is wound in the opposite sense, so the fields from the two coils cancel at the vapour sensor. This reduces direct RF pickup while the sample's oscillating nuclear magnetisation produces the signal measured by the atomic sensor.

Vapour cell, oven, and cooling: the 2 cm cubic cell contains potassium vapour, with nitrogen and helium buffer gases. The oven controls the cell temperature and potassium density; the cooling assembly helps manage heat around the RF coils and sample region.

Detuning coils: these coils shift the atomic resonance during the strong sample-excitation pulse, helping protect the potassium spin polarisation from the pulse.

Pump Laser And Optical Readout

Tapered amplifier: it boosts the pump-laser power so the expanded beam can illuminate the large vapour cell with sufficient, relatively uniform optical power. The acousto-optic modulator (AOM) provides fast control of the pump beam, allowing it to be switched or modulated in sync with the measurement sequence.

Field Control And Readout

XYZ coils: arranged in Helmholtz geometry, these coils provide the static tuning field B0 that brings the atomic Larmor frequency into resonance with the NQR signal. They also compensate residual static magnetic fields. The separate RF-drive coil supplies a smaller oscillating field for characterising or driving the sensor response.

The pump and probe optics, balanced photodetector, and lock-in readout follow the same general arrangement as the QEP 2.0 Atomic Magnetometer. Here, the probe rotation carries the response to the sample's oscillating nuclear magnetisation. The RF amplifier and matching network drive the sample excitation coils, while the function generator and data-acquisition system coordinate the excitation and measurement timing.

Citations

S.-K. Lee, K. L. Sauer, S. J. Seltzer, O. Alem, and M. V. Romalis, “Subfemtotesla radio-frequency atomic magnetometer for detection of nuclear quadrupole resonance,” Applied Physics Letters, vol. 89, 214106 (2006).

A. N. Garroway, M. L. Buess, J. P. Yesinowski, and J. B. Miller, “Narcotics and explosives detection by 14N pure NQR,” SPIE Substance Detection Systems, vol. 2092, p. 318 (1993).

Optics And Engineering

Selected apparatus images showing the coil assembly and experimental hardware.

CAD rendering of the NQR magnetometer assembly with large blue circular coils around the vapour cell and sample region
CAD rendering of the coil-and-cell assembly, with the blue rings arranged around the vapour cell and sample region.
Three-axis Helmholtz coils mounted around a removable inner cage
Three-axis Helmholtz coils surround the inner cage, designed to slide in and out of the magnetic and RF shields.
Square 20 by 20 by 20 potassium vapour cell with potassium droplets visible inside
The 20 × 20 × 20 vapour cell, with potassium droplets visible inside.
Assembled NQR magnetometer lantern structure mounting the detuning coils, RF coils, cooling, oven, and vapour cell
The assembled lantern structure mounts the detuning and RF coils, cooling and oven systems, and vapour cell.